產品編號: 58044001
| 氯化物測量範圍 | 0-300 ppb |
| 氯化物偵測限制 | 0.5 ppb |
| 氯化物測量精度 | 讀數的±5%或± 0.5 ppb; 以較高者為準 |
| 硫酸鹽測量範圍 | 0-300 ppb |
| 硫酸鹽偵測限制 | 1 ppb |
| 硫酸鹽測量準確度 | |
| 測量週期時間 | 一般45分鐘,可設定15分鐘與1小時 |
| 樣品流動速率 | 25至50mL/min |
| 樣品溫度 | 10-45°C (50-113°F) |
| 樣品壓力 | 0.3至7bar(5至100psig) |
| 採樣測量 | 隨附; 100mL容量 |
| 電源 | 100-240 VAC,50-60 Hz,一般為100 W |
| 尺寸(HxWxD) | 927x508x305 mm (36.5"x20.9"x12") |
| 重量 | 44公斤(97 磅) |
| 環境操作溫度 | 10-35°C (50-95°F) |
| 濕度 | 10-70%,非冷凝 |
| 等級/認證 | CE / cULus |
| 製程連接 | 樣品入口: 1/4"或6 mm OD管SS壓合接頭; 排水軟管: 內含19x25.4 mm (3/4 x 1") 2 m (6 ft) 長度 |
| 操作員介面 | 彩色觸控螢幕 同步顯示離子濃度與分析儀狀態 |
| 類比輸出 | 22mA警報,500歐姆最大負載,不可與外部供電的電路使用 |
| 類比輸出調節 | 線性、雙線性、對數式(1、2、3、4 個對數尺度),自動設定範圍 |
| 測量參數 | 氯化物和硫酸鹽 |
| 繼電器接點 (2) | 機械等級250 VAC,3 Amps(繼電器1為 NC;繼電器2至4為NO),4-SPDT 類型Reed 250 VAC或DC,0.5 Amps(繼電器5至8) |
| 類比輸出精度 | ±0.05 mA |
| 型號系列 | 3000CS |
| 外殼類型 | 完整 |
| 內建定序儀 | 否 |
| Versatile for Critical Applications The 3000CS is designed for measurement in a range of power plant water applications, including steam quality monitoring at the turbine inlet, condensate monitoring at the condensate polisher, boiler feedwater monitoring, and makeup water quality monitoring. | ![]() |
| Intuitive Interface with Data Trends This on-line chloride and sulfate analyzer uses the METTLER TOLEDO M800 transmitter with intuitive touchscreen interface to easily monitor cycle chemistry. It offers unique information such as trendlines for each measurement. | ![]() |
| Easy Integration for Data Management The 3000CS has an analog output for record keeping of chloride and sulfate contamination levels in your data acquisition software. This helps ensure that you stay within the specs of your turbine warranty, and helps you identify causes of contamination. | ![]() |
| Durable Enclosure for Long Lifetime A full enclosure protects the reagent containers and components of the 3000CS from harsh environmental conditions, ensuring a measurement uneffected by outside influences. The 3000CS operates in humidities from 10% to 70% (non-condensing). | ![]() |