The new revision of the mandatory Chapter 41 introduces the following updates starting February 1, 2026:
- Calibration must be performed periodically, including measurement uncertainty, with frequency based on risk.
- Calibration is required before and after any operation that may affect balance performance (as found and as left).
- Periodic risk-based performance checks on sensitivity and repeatability must be conducted between calibrations.
- EURAMET cg-18 and ASTM E898 are mentioned as applicable standards for balance calibration