Material No.: 58044001
| Chloride Measurement Range | 0-300 ppb |
| Chloride Limit of Detection | 0.5 ppb |
| Chloride Measurement Accuracy | ±5% of reading or ± 0.5 ppb whichever is greater |
| Sulfate Measurement Range | 0-300 ppb |
| Sulfate Limit of Detection | 1 ppb |
| Sulfate Measurement Accuracy | ±10% of reading or ±1 ppb whichever is greater |
| Measurement Cycle Time | 45 min typical, programmable between 15 minutes and 1 hour |
| Sample Flow rate | 25-50 mL/min |
| Sample Temperature | 10-45°C (50-113°F) |
| Sample Pressure | 0.3-7 bar (5-100 psig) |
| Grab Sample Measurement | Included 100 mL capacity |
| Power Supply | 100-240 VAC, 50-60 Hz, 100 W typical |
| Dimensions (HxWxD) | 927x508x305 mm (36.5"x20.9"x12") |
| Weight | 44 kg (97 lbs) |
| Ambient Operating Temperature | 10-35°C (50-95°F) |
| Humidity | 10-70% non-condensing |
| Ratings/Approvals | CE / cULus |
| Process Connection | Sample Inlet: 1/4" or 6 mm OD tube SS compressin fitting Drain Hose: 19x25.4 mm (3/4 x 1") 2 m (6 ft) length included |
| Operator Interface | Color touchscreen Simultaneous display of ion concentrations and analyzer status |
| Analog Output | 22mA alarm, 500 ohm max load, not for use with externally powered circuit |
| Analog Output Scaling | Linear, bi-linear, logarithmic (1,2,3,4 decades), auto ranging |
| Measurement Parameter | Chloride and sulfate |
| Relay Contacts | Mechanical rated at 250 VAC, 3 Amps (Relay 1 NC, Relay 2 to 4 NO), 4-SPDT Type Reed 250 VAC or DC, 0.5 Amps (Relay 5 to 8) |
| Analog Output Accuracy | ±0.05 mA |
| Model Family | 3000CS |
| EnclosureType | Full |
| Built-in sequencer | No |