Material No.: 30039086
| Process Connection | ¼" NPT(F) in/ out |
| Operating Temperature (Celsius) | 0 °C - 100 °C |
| Operating Temperature (Fahrenheit) | 32 °F - 212 °F |
| Sample Flow rate | 50 to 150 ml/min |
| Sample Conductivity > (μS/cm) | 0.3 |
| Refillable Electrolyte | 9823 3M KCl |
| pH Range | 1 – 12 |
| Sterilizable/Autoclavable | No |
| Sensor Length | 120 mm |
| Cable Connector | AK9 |
| Operating Pressure (Bar) | 0-7 barg @ 100 °C |
| Model Family | pHure |
| Other certificates | CE |
| pH Sensing Technology | Steam Sterilizable Glass Membrane (A41) |
| Reference System | 3M KCl |
| Temperature Sensor | Digital |
| Hazardous Area Approvals | No |
| ISM (Digital) or Analog | ISM (Digital) |
| Measurement Parameter | ORP/Redox pH |
| Primary Media/Application | Other Water Processes Power Plant Water/Steam Semiconductor Water |
| Shaft Material | Glass |
| Hygienic/Biocompatibility Approvals | No |
| Operating Pressure (PSI) | 0-101 psig @ 212 °F |
| Solution Ground/Redox Metal | Platinum |