Material No.: 52206437
| Measurement Parameter | Gas-Phase Oxygen |
| Measurement Technology | Amperometric Clark Electrode |
| Measurement Range | 51 ppm - 60 V-% O2 |
| Accuracy | ≤ ± [1% +50 ppm] |
| Response Time | T90 ≤ 20s (nitrogen to 15 Vol-% O2 at 25 °C / 77 °F) |
| Primary Media/Application | Chemical Processes Pharmaceutical/Life Science Production |
| ISM (Digital) or Analog | ISM (Digital) |
| Cable Connector | K8S |
| Sensor Length | 120 mm |
| Process Connection | Pg 13.5 |
| Connector Design | Straight |
| O-Ring Material | Silicone |
| Sensor Diameter | 12 mm |
| Surface Roughness of Wetted Parts | N5/Ra 16 (Ra=0.4mm/16min) |
| Wetted Materials | 316L Stainless Steel |
| Wetted membrane material | PTFE |
| Model Family | InPro6900iG |
| Detection limit | 50 ppm |
| Detection limit | 50 ppm |
| Sensor signal in air at 25 °C (77 °F) | 250...500 nA |
| Residual signal in oxygen free media | <0.03 % of signal in ambient air |
| Temperature compensation | Automatic |
| Measuring Temp. Range | 32 °F – 158 °F (0 °C – 70 °C) |
| Operating Pressure (Bar) | 0.200 bar–9.000 bar |
| Operating Pressure (PSI) | 2.9 to 130 psi absolute |
| Mechanical Temp. Range | 23 °F – 250 °F (-5 °C – 121 °C) |
| Design pressure (bar) | maximum 12 bar |
| Design Pressure (PSI) | maximum 174 psi absolute |
| Hazardous Area Approvals | ATEX: Ex ia IIC T6 / T5 / T4 / T3 Ga / Gb, Ex ia IIIC T69 °C / T81 °C / T109 °C / T161 °C Da / Db. FM: IS Cl. I, II, III, Div 1, GR ABCDEFG / T6 * |
| Hygienic/Biocompatibility Approvals | EHEDG, FDA/USP Class VI |
| Other certificates | METTLER TOLEDO Quality Certificate, Material certificate 3.1, N5/Ra 17 |
| Mechanical Temperature Resistance (Celsius) | -5 °C – 121 °C |
| Operating Range | 51 ppm - 60 V-% O2 |
| Connection | K8S ISM |
| Measured Pressure Resistance | 0.2 bar – 9 bar |
| Mechanical Temperature Resistance (Fahrenheit) | 23 °F – 250 °F |