Material No.: 58045001
| Operating Range (Temp.) | 0 °C – 37 °C |
| Operating Range (Humidity) | 10 % – 90 % |
| Flow Rate | 30 mL/min – 100 mL/min |
| Dimensions (HxWxD) | 616 mm x 564 mm x 305 mm |
| Analysis Time | Continuous |
| Wall Mount | Anti-vibration shelf required (P/N 58 079 700) |
| Minimum detection size | ≥ 0.3 μm |
| Inlet pressure | 21 - 80 psig (2 - 5.5 bar(g)) Calibration, cleaning and grab sample requires sample pressure of 0 psig (0 bar(g) Process pressure above 80 psig (5.5 bar(g)) requires optional High Pressure Regulator (P/N 58 091 552) |
| Humidity (non-condensing) | 80% maximum relative humidity up to 31 °C (87.8 °F) decreasing linearly to 50% relative humidity at 40 °C (104 °F) |
| Ambient temperature (non-condensing) | 1 - 37 °C (32 - 98.6 °F) Temperature below 15 °C or above 45 °C requires Sample Conditioning Coil (included) |
| Voltage | MAINS supply voltage fluctuations up to ±10% of the nominal voltage of 100 - 240 VAC 50 - 60 Hz TEMPORARY OVERVOLTAGES occurring on the MAINS SUPPLY TRANSIENT OVERVOLTAGES: up to levels of OVERVOLTAGE CATEGORY II |
| Data communication | 5 - 20 mA standard, with configurable output ranges Analog output channels Ethernet - standard RJ 45 / Wi-Fi capable SCADA connectivity via Modbus TCP USB |
| Enclosure Material | Stainless Steel |
| Measurement Range | 1 -10,000 AFU / mL |
| Response Time | 3 seconds (1 mL) |
| Physical dimensions | (WxHxD) 22.2"(56.4 cm) W x 24.25"(61.6 cm) H x 12"(30.5 cm) D |
| Monitoring location | At-line to drain |
| Flow rate | 31 mL/min |
| Use | Indoor use |
| Environmental Temperature | 6 - 35 °C (41- 95 °F) |
| Weight | 73.4 lbs (33.3 kg) |
| Outlet connection | 0.125" (3 mm) O.D. |
| Inlet connection | 0.125" (3 mm) O.D. |
| Power Requirements | 100 - 240 VAC 50 - 60 Hz 5A Use the power cord included with the instrument 8.2' (2.5 m) cord length provided standard |
| Temperature (non-condensing) | 6 - 90 °C (41-194 °F) Temperature below 15 °C or above 45 °C requires Sample Conditioning Coil (included) |
| Environment | Pollution degree 3 |
| Detection limit | 2 AFU (Auto Fluorescent Units) |
| Type / Quality | Purified Water (PW), Ultrapure Water (UPW), Water for Injection (WFI) |
| Altitude | Up to 6562' (2000 m) |